論文

査読有り
2011年

Measurement of absolute detection efficiencies of a microchannel plate using the charge transfer reaction

Physica Scripta T
  • N. Takahashi
  • ,
  • S. Hosokawa
  • ,
  • M. Saito
  • ,
  • Y. Haruyama

T144
記述言語
英語
掲載種別
研究論文(国際会議プロシーディングス)
DOI
10.1088/0031-8949/2011/T144/014057

A new method for determining the absolute detection efficiencies of a microchannel plate (MCP) detector is presented. This method uses a charge transfer reaction and can be applied to determine not only the detection efficiencies for ions but also those for neutral atoms. The measurements were carried out for Ar+ and neutral rare gas atoms (Ne, Ar, Kr and Xe) in the 0.5-4.8 keV energy range. The detection efficiencies for all species increase with increasing impact energy and approach the open area ratio of the MCP used (about 50%). The measured detection efficiencies were found to scale with the Lindhard, Scharff and Schiott formula for electronic stopping power at keV energies. © 2011 The Royal Swedish Academy of Sciences.

リンク情報
DOI
https://doi.org/10.1088/0031-8949/2011/T144/014057
ID情報
  • DOI : 10.1088/0031-8949/2011/T144/014057
  • ISSN : 0281-1847
  • SCOPUS ID : 80051704344

エクスポート
BibTeX RIS