Signal pulse emulation for scintillation detectors using Geant4 Monte Carlo with light tracking simulation

REVIEW OF SCIENTIFIC INSTRUMENTS
  • R. Ogawara
  • ,
  • M. Ishikawa

87
7
記述言語
英語
掲載種別
DOI
10.1063/1.4959186
出版者・発行元
AMER INST PHYSICS

The anode pulse of a photomultiplier tube (PMT) coupled with a scintillator is used for pulse shape discrimination (PSD) analysis. We have developed a novel emulation technique for the PMT anode pulse based on optical photon transport and a PMT response function. The photon transport was calculated using Geant4 Monte Carlo code and the response function with a BC408 organic scintillator. The obtained percentage RMS value of the difference between the measured and simulated pulse with suitable scintillation properties using GSO:Ce (0.4, 1.0, 1.5 mol%), LaBr3:Ce and BGO scintillators were 2.41%, 2.58%, 2.16%, 2.01%, and 3.32%, respectively. The proposed technique demonstrates high reproducibility of the measured pulse and can be applied to simulation studies of various radiation measurements. (C) 2016 Author(s).