論文

査読有り
2006年12月

Performance characterization of recombinant L-glutamate oxidase in a micro GOT/GPT sensing system

SENSORS AND ACTUATORS B-CHEMICAL
  • Sanjay Upadhyay
  • ,
  • Naoto Ohgami
  • ,
  • Hitoshi Kusakabe
  • ,
  • Hiroshi Mizuno
  • ,
  • Jiro Arima
  • ,
  • Takashi Tamura
  • ,
  • Kenji Inagaki
  • ,
  • Hiroaki Suzuki

119
2
開始ページ
570
終了ページ
576
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1016/j.snb.2006.01.008
出版者・発行元
ELSEVIER SCIENCE SA

Large-scale production Of L-glutamate oxidase (L-GlOx) from Streptomyces sp. X-119-6 was conducted with the use of an Escherichia coli expression system. The active alpha(2)beta(2)gamma(2)-subunit structure of the enzyme was obtained by proteolysis of a precursor form using metalloendopeptidase. The performance of this recombinant enzyme was tested in an amperometric sensing system to determine the L-glutamate concentration. To this end, a thin-film, three-electrode system was formed. Hydrogen peroxide produced by an enzymatic reaction was detected by using a platinum working electrode. The sensitivity of the L-glutamate sensor was 220 nA/mM, and the lower detection limit was 3 mu M (S/N = 3). Up to 800 mu M, a linear relationship was observed between the output current and the L-glutamate concentration. Furthermore, we used a microanalysis system to determine the activities of glutamic oxaloacetic transaminase (GOT) and glutamic pyruvic transaminase (GPT) and, ultimately, to test whether the recombinant enzyme can additionally be used to analyze enzyme activities. A polydimethylsiloxane (PDMS) flow channel with a mixing compartment was also used. The substrate solution and the sample solution for either GOT or GPT were mixed as they were being removed from the flow channel. As time elapsed, the L-glutamate concentration in the mixed solution increased because of the enzymatic reaction of GOT or GPT; consequently, a constant increase in current was observed. The relation between the slope of the response curve and the enzyme activity was linear up to 127 U/l for GOT and to 88 U/l for GPT. The results of this series of experiments demonstrate that recombinant L-GlOx can be used to construct micro sensors and microanalysis systems. (c) 2006 Elsevier B.V. All rights reserved.

リンク情報
DOI
https://doi.org/10.1016/j.snb.2006.01.008
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000241197600032&DestApp=WOS_CPL
ID情報
  • DOI : 10.1016/j.snb.2006.01.008
  • ISSN : 0925-4005
  • Web of Science ID : WOS:000241197600032

エクスポート
BibTeX RIS