Evaluation of portable blood glucose meters using canine and feline pooled blood samples

POLISH JOURNAL OF VETERINARY SCIENCES
  • A. Mori
  • ,
  • H. Oda
  • ,
  • E. Onozawa
  • ,
  • S. Shono
  • ,
  • T. Takahashi
  • ,
  • S. Yamashita
  • ,
  • H. Fujimoto
  • ,
  • T. Sako

19
4
開始ページ
707
終了ページ
713
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1515/pjvs-2016-0089
出版者・発行元
POLSKA AKAD NAUK, POLISH ACAD SCIENCES, UNIV WARMIA & MAZURY OLSZTYN

This study evaluated the accuracy and reproducibility of a human portable blood glucose meter (PBGM) for canine and feline whole blood. Reference plasma glucose values (RPGV) were concurrently measured using glucose oxidation methods. Fifteen healthy dogs and 6 healthy cats were used for blood sampling. Blood glucose concentrations and hematocrits were adjusted using pooled blood samples for our targeted values. A positive correlation between the PBGM and RPGV was found for both dogs (y = 0.877, x = -24.38, r = 0.9982, n = 73) and cats (y = 1.048, x = -27.06, r = 0.9984, n = 69). Acceptable results were obtained in error grid analysis between PBGM and RPGV in both dogs and cats; 100% of these results were within zones A and B. Following ISO recommendations, a PBGM is considered accurate if 95% of the measurements are within +/- 15 mg/dl of the RPGV when the glucose concentration is <100 mg/dl and within +/- 15% when it is >= 100 mg/dl; however, small numbers of samples were observed inside the acceptable limits for both dogs (11%, 8 of 73 dogs) and cats (39%, 27 of 69 cats). Blood samples with high hematocrits induced lower whole blood glucose values measured by the PBGM than RPGV under hypoglycemic, normoglycemic, and hyperglycemic conditions in both dogs and cats. Therefore, this device is not clinically useful in dogs and cats. New PBGMs which automatically compensate for the hematocrit should be developed in veterinary practice.

リンク情報
DOI
https://doi.org/10.1515/pjvs-2016-0089
PubMed
https://www.ncbi.nlm.nih.gov/pubmed/28092618
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000391349900004&DestApp=WOS_CPL
URL
https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85010424289&origin=inward
ID情報
  • DOI : 10.1515/pjvs-2016-0089
  • ISSN : 1505-1773
  • eISSN : 2300-2557
  • PubMed ID : 28092618
  • SCOPUS ID : 85010424289
  • Web of Science ID : WOS:000391349900004

エクスポート
BibTeX RIS